O7E Multi-Cell Ageing Oven | Wallace Instruments
Wallace Instruments - rubber testing products

LI-IN-Bug blue     X%20Logo   

O7E Multi-Cell Ageing Oven

O7E Multicell Ageing Oven

Accelerated ageing and heat resistance tests are designed to estimate the relative resistance of rubber to deterioration with the passage of time. The Wallace O7E Multi-Cell Ageing Chamber, allows samples to be aged in separate cells, preventing any contamination from adjacent samples through the migration of volatile substances.

Download the datasheet

 

Features

· Stable temperature range 60 - 250°C

· Seven separate test cells

· No volatile migration between cells

· Auto-tuning temperature control

· Adjustable air flow

· Accurate temperature and airflow

· Safety Temperature cut-out

· Sturdy Construction on stand and castor wheels

 

Principle of Operation

 

The rubber is subjected to controlled deteriorating influences for definite periods, after which appropriate properties are measured and compared with the corresponding properties of the unaged rubber.

This air oven method of ageing exposes a test sample to elevated air temperatures, allowing its physical properties, such as tensile strength, to be measured and compared with those of an unaged sample.

Avoiding the need for valuable bench space, the O7E is designed as a complete unit with its own stand and castor wheels.

The oven is a solid aluminium block with seven vertical cells machined from it.  Preheated air at a controlled temperature is pumped from a manifold through a calibrated opening into the bottom of each cell, ensuring even air and temperature distribution. 

To avoid contamination from re-circulation, the air is discharged to the atmosphere through the two covers at the top of each cell. The test samples are suspended from the inside of the top cover.

The oven includes a PID temperature controller with digital display. The controller is auto-tuning, achieving a high level of accuracy and stability at the working temperature. Alarms on the controller abort the tests if the temperature goes outside its operating band. 

A valve positioned on the top of the control module regulates the airflow, which is measured accurately by a calibrated
flow-meter.

A USB communication option is offered, allowing connection to a PC.

 

Safety Features

 

· After a power failure the oven will not auto-reset for safety reasons.

· The oven is fitted with thermostats, which will switch off the power supply if the cell temperature exceeds 280°C.

 

Learn More: UKAS 17025 Calibration

View and send my quotation »